Strain Gauged Column accessories
Strain Gauged Column
Strain Gauged Column 2MN for EN 12390-4
Dimensions (WxDxH)130x108x58 mm
Connections4 (One for each strain-gauge)
Sensor excitation5 V
Loadcell drive capacity150-2000 Ω
Sensor wiring4 or 6
Sensor input rangeTypical ±1-4 mV/V (adjustable 1-1000 mV/V) (Gain = 1 to 128)
Signa filtersinc3 or sinc4 filter: 1,2-1200 Hz Optional extra software filter in TramSmart (moving average)
Resolutionmax. ±2 million increments (±250,000 at 2 mV/V)
Temperature effect on zeroBetter than 1ppm/°C
Temperature effect on spanTypical 1ppm/°C
Store temperature range-20...+60°C (avoid wet or moist conditions)
Operating temperature rangeCompensated 0...+40°C (avoid wet or moist conditions)
DriftTypical 10 ppm /1000hours (gain 1)
Ingress ProtectionIP50
Weight0.5 kg